To evaluate the 3D static articulation accuracy of 3 model scanner-CAD systems (Ceramill Map400 [AG], inEos X5 [SIR], Scanner S600 Arti [ZKN]) using a coordinate measuring machine (CMM). Trueness and Show more
To evaluate the 3D static articulation accuracy of 3 model scanner-CAD systems (Ceramill Map400 [AG], inEos X5 [SIR], Scanner S600 Arti [ZKN]) using a coordinate measuring machine (CMM). Trueness and precision for each system will be reported in Part I. The master model simulated a single crown opposing a 3-unit fixed dental prosthesis. Five mounted stone cast sets were prepared, and one set was randomly selected. Reference values were obtained by measuring interarch and interocclusal reference features with the CMM. The stone cast set was scanned 5 times consecutively and articulated virtually with each system (3 test groups, n = 5). STL files of the virtual models were measured with CMM software. dR For trueness values, mean interarch global distortions ranged from 13.1 to 40.3 μm for dR The overall interarch global distortion of all three model scanner-CAD systems was low and did not exceed 0.6%. Variations in scanner technology, virtual articulation algorithm, and use of physical articulators contributed to the differences in distortion observed among all three groups. Show less
Accurate maxillomandibular relationship transfer is important for CAD/CAM prostheses. This study compared the 3D-accuracy of virtual model static articulation in three laboratory scanner-CAD systems ( Show more
Accurate maxillomandibular relationship transfer is important for CAD/CAM prostheses. This study compared the 3D-accuracy of virtual model static articulation in three laboratory scanner-CAD systems (Ceramill Map400 [AG], inEos X5 [SIR], Scanner S600 Arti [ZKN]) using two virtual articulation methods: mounted models (MO), interocclusal record (IR). The master model simulated a single crown opposing a 3-unit fixed partial denture. Reference values were obtained by measuring interarch and interocclusal reference features with a coordinate measuring machine (CMM). MO group stone casts were articulator-mounted with acrylic resin bite registrations while IR group casts were hand-articulated with poly(vinyl siloxane) bite registrations. Five test model sets were scanned and articulated virtually with each system (6 test groups, 15 data sets). STL files of the virtual models were measured with CMM software. dR Mean interarch 3D distortion ranged from -348.7 to 192.2 μm for dR Interarch and interocclusal distances increased in MO groups, while they decreased in IR groups. AG-IR had the greatest interarch distortion as well as interocclusal superior-inferior distortion. The other groups performed similarly to each other, and the overall interarch distortion did not exceed 0.7%. In these systems and articulation methods, interocclusal distortions may result in hyper- or infra-occluded prostheses. Show less
Prior studies have defined the accuracy of intraoral scanner (IOS) systems but the accuracy of the digital static interocclusal registration function of these systems has not been reported. This study Show more
Prior studies have defined the accuracy of intraoral scanner (IOS) systems but the accuracy of the digital static interocclusal registration function of these systems has not been reported. This study compared the three-dimensional (3D) accuracy of the digital static interocclusal registration of 3 IOS systems using the buccal bite scan function. Three IOS systems compared were 3M Mean 3D distortion ranged from -471.9 to 31.7 μm for dR Interarch and global interocclusal distortions for the three IOS systems were significantly different. TRC performed overall the best and TDS was the worst. The interarch (dR Show less